Epi Pen Recall

Mon, 04/03/2017 - 12:12pm

Message from Health Services:

Recently there was a national recall of Epipens in the United States.  Please see the press release below from Mylan.  The Health Services staff are currently identifying any assigned epipens in the health room for the affected lot numbers.  Parents with affected Epipens will be contacted for a replacement.  We WILL NOT discard of the Epipen while waiting for a replacement to be brought into school.  The affected Epipens will be returned to the families.  HCPSS Health Services is in contact with Mylan to obtain any necessary replacement for our STOCK Epipens as well.  Please feel free to contact Mrs. Wilson, NES Cluster Nurse, with any questions at amy_wilson@hcpss.org.

 

Press Release

This recall is being conducted as a result of the receipt of two previously disclosed reports outside of the U.S. of failure to activate the device due to a potential defect in a supplier component. The potential defect could make the device difficult to activate in an emergency (failure to activate or increased force needed to activate) and have significant health consequences for a patient experiencing a life-threatening allergic reaction (anaphylaxis). Both reports are related to the single lot that was previously recalled. The incidence of the defect is extremely rare and testing and analysis across the potentially impacted lots has not identified any units with a defect. However, the recall is being expanded to include additional lots as a precautionary measure out of an abundance of caution.
The recalled product was manufactured by Meridian Medical Technologies, a Pfizer company, and distributed by Mylan Specialty between December 2015 and July 2016. The expanded voluntary recall is being initiated in the U.S. and also will extend to additional markets in Europe, Asia, North and South America.
The recall impacts the 0.3 mg and 0.15 mg strengths of EpiPen Auto-Injector. None of the recalled lots include the authorized generic for EpiPen Auto-Injector, which is also manufactured by Meridian Medical Technologies. 
Mylan is committed to replacing recalled devices at no cost and Mylan would like to reassure patients that there will be no additional replacement-related financial burden to them as a result of this recall. Patients, customers and distributors are being notified and should refer to Mylan.com/EpiPenRecall for updates on product return and replacement instructions. We are asking patients to keep their existing product until their replacement product can be secured.
Patients may receive either EpiPen Auto-Injector or the authorized generic for EpiPen Auto-Injector at the pharmacy as a replacement based on availability. The authorized generic has the exact same drug formulation, has the exact same operating instructions and is therapeutically equivalent to EpiPen Auto Injector, and may be substituted for EpiPen Auto Injector.
It is important that patients continue to carry their current EpiPen Auto-Injector until they receive a replacement device. 
To return your product please contact Stericycle at 877-650-3494. If you have any additional questions regarding this recall, please contact Mylan Customer Relations at 800-796-9526 or customer.service@mylan.com.
Adverse reactions or quality problems experienced with the use of this product may be reported to the FDA's MedWatch Adverse Event Reporting program either online, by regular mail or by fax.
This voluntary recall is being conducted with the knowledge of FDA.
Epinephrine is the first-line treatment for a life-threatening allergic reaction (anaphylaxis) and access to this product is critical in the event of an emergency. Delays in epinephrine administration have been associated with negative health consequences. 
More information about the risks and benefits of EpiPen® Auto-Injector can be found at EpiPen.com.